Replacing the Display

The cover and fron t pan el must be r emoved befor e you can remove the
display. To remove it:
1
2
3
Step 3
Agilent L4532A/ 34A Service Guide
Remove all cables from the exter ior of the in str umen t, and remove the
cover (see
Removing the Cover
Remove the fron t pan el (see
In
st ep 6
of t hat pr ocedur e you discon n ect t he display cable fr om t he d isplay,
so t he cable is n ot shown below.
Remove the two # 10 Tor x round head scr ews holdin g the display
br acket to the front pan el.
on page 82).
Removing the Front Panel
Connector for Display Cable
5
Service and Repair
on page 92).
Step 3
95