Calibration and Adjustment

Agilent T echnologies Calibration Services

Agilen t Technologies offer s calibr ation ser vices at competitive prices.
When your digitizer is due for calibr ation , cont act your local Agilen t
Ser vice Cen ter for recalibration . See
page 72 for infor mation on contactin g Agilent.

Calibration Interval

The digitizer should be calibrated on a regular in ter val deter min ed by the
measuremen t accur acy requir ements of your application.
A 1 year in terval is adequate for most applications. Accur acy
specifications will r emain valid on ly if adjustmen t is made at r egular
calibr ation inter vals. Accuracy specifications are n ot valid beyond the
1 year calibr ation inter val. Agilent does not r ecommend extending
calibr ation inter vals beyon d 2 year s for any application .

Adjustment is Recommended

Specification s ar e on ly valid within the per iod stated from the last
adjustmen t. Whatever calibr ation in ter val you select, Agilent recommends
that complete re adjustment always be per for med at the calibr ation
in terval. This is n ecessar y to en sure that the digitizer will r emain within
specification for the next calibr ation interval. This cr iter ia for
r e adjustment provides the best lon g ter m stability.
Per for mance dat a measur ed during Per for mance Ver ification Tests does
n ot mean that the digitizer will remain within these limits un less the
adjustmen ts ar e per for med.

Time Required for Calibration

The digitizer can be automatically calibrated under computer control.
With computer control you can per for m the ver ification procedure in less
than 45 min utes and the adjustment procedur e in less than 10 min utes.
En sur e the digitizer is war med up before begin ning any of these
procedur es (see
44
Test Con sider ation s
Types of Ser vice Available
on page 48).
Agilent L4532A/ 34A Service Guide