Calibration and Adjustment
4
5

Procedure

1
2
3
4
5
60
En ter the Remote Control Page and select the
the followin g par ameter s an d click the
All Chann els:
Range ±250 mV
Coupling AC
Filter LP_20_MHz
Acquisition :
Sample Rate 20 kSa/ s
Recor ds 1
Samples/ r ecord 100000
Tr ig Holdoff 0 sec
Pr e- tr ig Samples/ r ecord 0
Tr ig delay 0
Tr igger :
Ar m source Immediate
Tr igger source Immediate
Select the
Acquire Data
Click the
a
View Measurements...
measuremen ts.
Click
an d select Vr ms for each digitizer chann el. Set the Ar ea
b
New
to Win dow. Set the Coupling to AC. Click the OK button .
Click the
button and set the St ar ting Offset to 0 and Window
c
Edit ...
Size to 100000. Click the OK button .
Leave the Measurements win dow open .
d
Set the Calibr ator to output 165 mVr ms at 1 kHz.
On the
tab, click the
Acquire Data
Note t he Vr ms measurements for the con nected chann el in the
a
In st alled Measur ements win dow.
Record the measur ed value as Vr ms
b
Use the
Configure Digitizer
the Apply Configuration button .
Set the Calibr ator output to 165 mVr ms at 1 MHz (43 Vr ms at 300 kHz
on the 64 V r ange).
On the
tab, click the
Acquire Data
Note t he Vr ms measurements for the con nected chann el in the
a
In st alled Measur ements win dow.
Record the measur ed value as Vr ms
b
Apply Configuration
tab.
button . Delete any installed
button.
Run Once
1kHz
tab to set the Sample Rate to 20 MSa/ s. Click
button.
Run Once
1MHz
tab. Set
Configure Digitizer
button.
.
.
Agilent L4532A/ 34A Service Guide