Procedure - Agilent Technologies L4532A Series Service Manual

20 msa/ s digitizer
Table of Contents

Advertisement

5

Procedure

1
2
3
N OT E
4
WA RNI NG
Agilent L4532A/ 34A Service Guide
Select the
Acquire Data
Click the
a
View Measurements...
measur emen ts.
Click
an d select Vave for each digitizer chann el. Set the Ar ea to
b
New
Window. Click the OK button .
Click the
button and set the Star tin g Offset to 0 an d Win dow
c
Edit ...
Size to 100000. Click the OK button.
Leave the Installed Measur emen ts win dow open .
d
Set the Calibr ator to output - 235 mVdc.
On the
tab, click the
Acquire Data
Note the Vave measur ements for all the chan nels in the In stalled
a
Measur ements window.
Compar e the measur ed values to the ones shown in the table below.
b
Set the Calibr ator to output each volt age in the followin g table. Use
the
t ab to chan ge the Filter or Range as in dicated in
Configure Digitizer
the table below.
The table uses the calibrator zero output for ease of connection. A more accurate
verification of the zero can be obtained by connecting a 50
input connectors.
If any values ar e out of r an ge, per for m the adjustments procedur e
(beginn ing on
page
64).
Shock Hazard. This procedure requires the application of up to 240
Vdc. Ensure all test connections are properly insulated and
shielded.
tab.
button . Delete any inst alled
button .
Run Once
Calibration and Adjustment
termination to the
4
53

Advertisement

Table of Contents
loading