Making 8510XF Adjustments

This chapter covers both the CRT and LCD displays. Some display-related
adjustments apply only to instruments with CRT display.
The following table provides adjustments most frequently needed for
8510XF system service. Most of the 8510XF is self-adjusting except for the
following:
Frequently used 8510XF adjustments
Adjustment Function
2
Aligns softkey labels and mechanical
softkey buttons.
Demagnetize the display.
Insures level power at the test ports
Assembly Adjusted
A11
Leveling DAC/ EEPROM, A16,
and Leveling Compensation
Board, A7
8510XF Service Quick Reference Guide 3-1