Parameter:
MIN = 1E-6 s
MAX = 10s
DEFault = 1s
1E-6 < value < 10
Initialization:
*RST
Example:
"SENSe:APERture
The aperture value is rounded to 1s
Interrogative form:
[:SENSe]:APERture? ==> <aperture time>
Returns the aperture time selected.
*************************************************************
[SENSe]:AVERage[:STATe]
Description:
Enables or disables the averaging of measurement results.
When the average function is enabled, the 2461 TYPE C takes
100 measurements and derives the average value. Note that
the resolution increases by one digit when this function is
enabled.
NOTE: This function is only available with frequency (on
channel 1 and channel 2), period, ratio and time interval
measurements.
Initialization:
*RST
Example:
"SENSe:AVERage
Enables averaging.
2461Type C User Manual
0.1s
0.6"
<ON | OFF>
OFF
ON"