Input 2 Sensitivity
Test
SIGNAL GENERATOR
AF OUT
Figure 6-2, Input 2 Low Frequency Sensitivity Test Configuration
Performance Verification and Adjustments 6-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Set the UCT to its default state. Select 50Ω input impedance for
input 2, and select external frequency standard.
Equipment: Signal Generator
Configure UCT as follows:
*RST
INP1:IMP 50
selects external frequency
If the UCT is fitted with a TCXO or OCXO oscillator option,
connect the UCT internal reference output to the external signal
source 10MHz Input. Configure the external signal source to
accept the 10MHz reference from the UCT.
Connect the Function Generator to the UCT as shown in Figure
6-2.
10MHz OUT
Apply the frequencies given in the following table to the UCT.
Check the UCT gives a stable reading within the tolerances
given, at less than the signal levels given.
Command Syntax:
MEAS1:FREQ?
2461Type C User Manual
returns unit to default state
selects 50Ω impedance
standard
IN
1
GATE
VT1
VT2
ARM
IN
EXTERNAL STANDARD INPUT
IN 3
STD
OUT
T2
IN
INPUT 2
2
OSC
UCT
takes a frequency measurement
Racal Instruments 1998