2461Type C User Manual
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
"CONFigure1:TOTalize"
Configures the 2461 TYPE C to totalize events occurring at
channel 1 input during the time window defined by signal on
channel 2.
NO INTERROGATIVE FORM
*******************************************************
CONFigure[1 | 2][:SCALar]:PHASe [<expected value>,
<resolution>]
CONFigure[1 | 2]:ARRay:PHASe <array size> [,<expected value>,
<resolution>]
DESCRIPTION:
These commands configure the 2461 TYPE C to measure
the phase difference between the waveform applied to the
channel 1 input and that applied to the channel 2 input
(Phase 1 relative to 2), or between the waveform applied to
the channel 2 input and that applied to the channel 1 input
(Phase 2 relative to 1).
Note: The phase measurement is aborted and an error is
generated if the frequencies of the signals applied to channel
1 and 2 are different.
The ARRay command programs the counter for multiple
phase measurements as specified by the <array size>
parameter.
The optional <expected value> and <resolution> parameters
are accepted but ignored. The units are in degrees. The
resolution is determined by the input signal.
INITIALIZATION:
*RST Resolution = 8
EXAMPLE:
"CONFigure2:PHASe 100"
Configures the 2461 TYPE C to measure the phase
difference between the waveform applied to the channel 2
input and that applied to channel 1 input.
Programming The Type C Function Card 4-21
Array size = 1