Interrogative form:
SENSe:TOTalize:GATE:STATe?
Returns the current state of the totalize function.
*******************************************************
[SENSe[1 | 2]]:TOTalize:GATe:POLarity <NORMal |
INVerted>
Description:
Selects the slope of the selected channel. NORM
selects a positive slope, and INV selects a
negative slope.
This function is only available if the currently selected
function is TOTalize. Sending this command while
another measurement function is selected generates the
error –240 ("Function not allowed").
Initialization:
*RST
Example:
CONF1:TOT
"SENSe2:TOTalize:GATE:POLarity INV"
Selects a negative polarity for channel 2 (the
gating channel).
Interrogative form:
SENSe1:TOTalize:GATe:POLarity? ==>
<NORMal | INVerted>
Returns the slope polarity of channel 1.
**********************************************************
[SENSe[1 | 2]]:TOTalize:INTermediate <ON | OFF | DEF>
DESCRIPTION:
Enables or disables intermediate (i.e. on the fly) totalize
measurements. If this feature is enabled, FETCh?
queries sent during the totalize measurement will return
the last intermediate result.
Programming The Type C Function Card 4-57
==>
NORM (both channel 1 and 2).
<ON | OFF>