Racal Instruments 2461 User Manual page 126

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
"Positive pulse width"
"Negative pulse width"
"Volts middle, Input 1"
"Volts middle, Input 2"
"Volts maximum, Input 1"
"Volts maximum, Input 2"
"Volts minimum, Input 1"
"Volts minimum, Input 2"
The Resolution text box controls the number of digits of
resolution of measurement results.
The Average binary switch enables or disables the averaging of
measurement results. When Average is enabled, the 2461 Type
C takes 100 measurements and calculates the average value.
The Input1 and Input2 text boxes in the Manual Trigger Level
group control the trigger levels of the two channels when
Manual Trigger is set. Note that updating these levels while in
Auto Trigger mode automatically enables Manual Trigger mode.
The Open Gate and Clear controls manage the Manual Totalize
function. The Open Gate command button is used to start and
stop the counting. Successive counts are cumulative. The Clear
button allows the user to reset the Totalize count.
The Single Shot and Continuous buttons control the actual
measurement process. Pressing the Single Shot button initiates
a 2461 Type C measurement and acquires the result. Pressing
the Continuous button causes the 2461 Type C to take
continuous measurements with the results reported in real-time.
The Abort button aborts the measurement in progress by
forcing the 2461 Type C into the Idle State.
The RESET button resets only the currently selected UCT if
there are two UCT's installed.
The Help button supports on-line help that provides the user
with an efficient way to use the panel without referring to printed
documentation.
The Arming button (Configuration group) manages the source
and slope selection of the arming circuits. These settings are
contained in the secondary panel illustrated in figure 5-2.
The Inputs button (Configuration group) displays the secondary
panel illustrated in figure 5-3 and allows the user to specify
some 2461 Type C input characteristics, such as attenuation,
coupling, impedance, etc.
The Frequency Standard button (Configuration group)
Using The VXI plug&play Install Disks 5-3