Totalize 1 by 2 Test

Performance Verification and Adjustments 6-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Set the UCT to its default state. Select Totalize 1 by 2 as the
measurement mode, 50
frequency standard.
Equipment: Signal Generator
Configure UCT as follows:
*RST
INP1:IMP 50
SOUR:ROSC:SOUR EXT
If the UCT is fitted with a TCXO or OCXO oscillator option,
connect the UCT internal reference output to the external signal
source 10MHz Input. Configure the external signal source to
accept the 10MHz reference from the UCT.
Connect the UCT to the signal generator as shown in Figure 6-
6. Set the signal generator to 10MHz, 0dBm. Check that the
UCT measurement result is 5000 +/- 1.
SIGNAL GENERATOR
RF OUT

Figure 6-4, Totalize 1 by 2 Test Configuration

Command Syntax for making a Totalize measurement:
CONF1:TOT
MEAS1:TOT?
input 1 impedance, and external
return unit to default state
selects 50Ω impedance for
Input 1
selects external frequency
standard
GATE
VT1
VT2
ARM
IN
STD
OUT
1KHz AF OUT
T2
10MHz OUT
UCT
configures UCT for Totalize
measurement
takes a totalize measurement
Racal Instruments  1998
2461Type C User Manual
IN
INPUT 1
1
EXTERNAL STANDARD INPUT
IN 3
IN
INPUT 2
2
OSC