Read Subsystem - Racal Instruments 2461 User Manual

200mhz universal counter timer type c
Table of Contents

Advertisement

READ Subsystem

COMMAND
READ?
Programming The Type C Function Card 4-38
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
READ command summary
Initiates a measurement and returns the value.
READ?
Description:
Causes an INITiate:IMMediate action followed by a FETCh?
query. INITiate:IMMediate performs the data acquisition, while
FETCh? returns the measurement.
The measurement function must be selected using the
CONFigure:FUNCtion, SENSe:FUNCtion or
MEASure:FUNCtion? commands.
The number of measurements taken and returned depend on
the <array size> parameter specified in the last
CONFigure:ARRay or MEASure:ARRay commands. Only one
measurement is taken if the last CONFigure or MEASure
command specified a SCALar-type measurement.
Note that, in the case of ARRay-type measurement, the time-out
period (refer to SYSTem:TIMeout) may have to be increased.
Initialization:
Not Applicable
Example:
READ?
Initiates and retrieves a measurement on the selected
function
2461Type C User Manual
DESCRIPTION
Racal Instruments  1998

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents