Connects or separates input channel 1 and 2. When
COMMon is specified, the signal at channel 1 is connected to
both channels. When this is done, the channel 2 input socket
is isolated from channel 2.
If COMMon is specified while the functions
VOLTage:MIDDle, VOLTage:MAXimum and
VOLTage:MINinimum are active, then the error –240 is
generated ("Function Error").
If SEParate is specified while the functions PHASe,
RISE:TIME, FALL:TIME, PWIDTH and NWIDTH are active,
then the error –240 is generated ("Function Error").
PARAMETERS:
DEFault = SEP
INITIALIZATION:
*RST SEP
INTERROGATIVE FORM
INPut[1 | 2]:ROUTe? ==> <SEParate | COMMon>
Returns the current input route setting.
If the functions VOLTage:MIDDle, VOLTage:MAXimum and
VOLTage:MINinimum are active, then the query returns
SEParate.
If the functions PHASe, RISE:TIME, FALL:TIME, PWIDTH
and NWIDTH are active, then the query returns COMMon.
*******************************************************
INPut[1 | 2]:FILTer[:LPASs][:STATe] <ON | OFF | DEFault>
DESCRIPTION:
Selects the input block signal low-pass filter state for the
specified channel to ON or OFF.
PARAMETERS:
DEFault = OFF