TRIGGER SUBSYSTEM
INITiate
ABORt
ARMing

Functional Check

INTRODUCTION

VXIBUS
Communication
Check
Programming The Type C Function Card 4-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
COMMAND
:(CONTinuous | [IMMediate])
:([SEQuence1] | STARt)[:LAYer]
:SLOPe
:SOURce
:(SEQuence2 | STOP)
:SLOPe
:SOURce
The following procedure verifies that the 2461 TYPE C can accept,
process, and transmit device- dependent messages over the
VXIbus.
Perform the following procedure:
(1)
Power-on the VXI Chassis/Mainframe containing the 2461
(2)
Perform a test as shown below:
Action
Send the IEEE-488.2
command "*TST?"
(3)
Read the data in the output buffer.
Action
Self Test Data
Read Data
"0" = Pass NON Zero value indicates a failure
2461Type C User Manual
*RST OR
SYSTEM:PRESET
OFF
POS
IMM
POS
IMM