EXAMPLE:
"CONFigure1:FALL:TIME"
Configures the 2461 TYPE C to measure the fall time of the sig
applied to channel 1.
NO INTERROGATIVE FORM
*******************************************************
CONFigure[1 | 2][:SCALAR]:PERiod [<EXPECTED VALUE>,
<RESOLUTION>]
CONFigure[1 | 2]:ARRay:PERiod <ARRAY SIZE>[,<EXPECTED
VALUE>, <RESOLUTION>]
DESCRIPTION:
These commands configure the 2461 TYPE C to measure
the period of the signal on the input channel selected.
The ARRay commands program the counter for multiple
period measurements as specified by the <array size>
parameter.
The optional <expected value> and <resolution> parameters
are used to derive the required number of digits of resolution.
Their units are seconds.
INITIALIZATION:
*RST Resolution = 8
EXAMPLE:
"CONFigure1:PERiod"
Configures the 2461 TYPE C to measure the period of the
signal applied to channel 1.
NO INTERROGATIVE FORM
*******************************************************
CONFigure[1][:SCALAR]:PWIDth [<EXPECTED VALUE>,
<RESOLUTION>]
CONFigure[1]:ARRay:PWIDth <ARRAY SIZE> [,<EXPECTED
VALUE>, <RESOLUTION>]
CONFigure[1][:SCALAR]:NWIDth [<EXPECTED VALUE>,
<RESOLUTION>]
CONFigure[1]:ARRay:NWIDth <ARRAY SIZE> [,<EXPECTED
VALUE>, <RESOLUTION>]
Programming The Type C Function Card 4-17
Array size = 1