Racal Instruments 2461 User Manual page 116

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description:
Selects the start arming source to be used when the 2461 TYPE
C is initiated.
When <CTR_EXTARM> is specified, then if the current source
for the stop arming is TTLTrg, this will be forced to
CTR_EXTARM.
When <TTLTrg> is specified, then if the current source for the
stop arming is CTR_EXTARM, this will be forced to TTLTrg.
If the TTLTrg line specified is already being used for the gate
output signal of one of the 2461 TYPE C boards in the system
(refer to the OUTPut subsystem), then the error –221 is
generated ("Setting conflict").
If the device is not in the Idle state this command has no effect
on the trigger system and an error –221 is generated ("Setting
conflict").
Initialization:
*RST
Example:
"ARMing:STARt:SOURce TTLTrg5"
Selects a TTL trigger line 5 for the start arming.
Interrogative form:
ARMing ([:SEQuence1] | :STARt)[:LAYer]:SOURce?
==> <CTR_EXTARM | TTLTrg<0..7> | IMMediate >
Returns the source of the start arming signal.
****************************************************************
ARMing:(SEQuence2 | STOP) [:LAYer]:SLOPe
<POSitive | NEGative | DEFault>
Programming The Type C Function Card 4-73
IMMediate

Hide quick links:

Advertisement

Table of Contents
loading