attenuation level, the following is set:
If channel 1 is specified, then manual trigger is set for
channel 1;
If channel 2 is selected and SEParate is specified, then
manual trigger is set for channel 2.
If channel 2 is selected and COMMon is specified, then
manual trigger is set for channel 1.
If this command is received while the functions PHASe,
RISE:TIME, FALL:TIME, PWIDTH and NWIDTH are active,
then the error –240 is generated ("Function Error").
If this command is received while the functions VOLT:MAX,
VOLT:MIN and VOLT:MIDD are active and the channel
specified is the same as the channel currently configured for
the measurement function, then the error –240 is generated
("Function Error").
PARAMETERS:
-5.1 ≤ value ≤ 5.1
DEFault = 0
MIN = -5.1
MAX = 5.1
INITIALIZATION:
*RST 0
INTERROGATIVE FORM
INPut[1 | 2]:COMParator:LEVel[:ABSolute]? ==>
<comparator level>
Returns the current comparator level setting for the specified
channel. Note that the attenuation setting is ignored.
INPut[1 | 2]:COMParator:HYSTeresis:RELative
<MINimum | MAXimum | DEFault>
Programming The Type C Function Card 4-45