The resolution of PHASE and TOTALIZE measurements is determined by the input signal.
Measurements of TIME INTERVAL, RATIO, PULSE WIDTH and RISE/FALLTIME have
resolution determined by both the input signal and the resolution set.
Note: a)To avoid unnecessary shifting of digits, the MSD is allowed to exceed the resolution
b) Measurement of FREQUENCY, PERIOD, RATIO and CHECK are averaged when
Note (ii) - Trigger Error
2
+
2
e
e
n
i
=
seconds RMS
Err
trig
S
trig
Where:
A-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selected number of
digits in Frequency,
(a)
Period and Check
10
9
8*
7
6
5
4
3
* initialized state
by 1 digit for a 10% over range.
these number of digits are selected.
Errtrig = Trigger Timing Error in seconds RMS.
= input amplifier RMS noise (typically 170 µV RMS in 200MHz bandwidth).
e
i
e
= input signal RMS noise in 200MHz bandwidth.
n
S
= slew rate of input signal at trigger point.