Racal Instruments 2461 User Manual page 84

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
*RST AC
INTERROGATIVE FORM
INPut[1 | 2]:COUPling? ==> <AC | DC>
Returns the current coupling setting for the specified
channel.
*******************************************************
INPut[1 | 2]:IMPedance <value | MINimum | MAXimum |
DEFault>
DESCRIPTION:
Selects the input terminating impedance for the specified
channel. If <value> is less than 60, the impedance is set to
50 Ω. If <value> is greater than 60, the impedance is set to
1M Ω.
PARAMETERS:
0 ≤ value ≤ 10E6Ω
MIN = 50Ω
MAX = 1E6Ω
DEFault = 1E6Ω
INITIALIZATION:
*RST
MAX
INTERROGATIVE FORM
INPut[1 | 2]:IMPedance? ==> <50 | 1000000>
Returns the current impedance setting for the specified
channel.
*******************************************************
INPut[1 | 2]:ROUTe <SEParate | COMMon | DEFault>
DESCRIPTION: