COMMAND
INITiate
:IMMediate
:CONTinous
[:STATe]
ABORt
ARMing
([:SEQuence1] | :STARt)[:LAYer]
:SLOPe
:SOURce
:(SEQuence2 | STOP) [:LAYer]
:SLOPe
:SOURce
Programming The Type C Function Card 4-70
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TRIGger command summary
Enter in the INITiate state.
Causes the 2461 TYPE C to operate in
continuous mode
Aborts and operation in progress
Slope of the arming signal for start.
Source of the arming signal for start.
Slope of the arming signal for stop.
Source of the arming signal for stop.
INITiate[:IMMediate]
Description:
The INITiate:IMMediate command causes the trigger system to
initiate and complete one full trigger cycle, returning to the Idle
state on completion.
If the device is not in the Idle state an IMMediate command has
no effect on the trigger system and an error –213 is generated
("Init ignored").
Initialization:
Trigger system in idle state.
Example:
"INITiate:IMMediate"
Initiates a single measurement.
*****************************************************
2461Type C User Manual