Racal Instruments 2461 User Manual page 165

200mhz universal counter timer type c
Table of Contents

Advertisement

×1 Attenuation:
×10 Attenuation:
Auto-Trigger Minimum Amplitude
150mV pk-pk.
Auto-Trigger Level Accuracy
(Relative to true trigger point)
×1
×10
Frequency 1 or 2
Range
Channel 1
6 x 10
Channel 2
6 x 10
LSD Displayed
-D
LSD = F x 10
Where:
Resolution
(selectable between 3 and 10 digits.)
For 3 to 5 Digits:
A-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selected if positive and negative peak amplitudes are less than
±4.6V and the peak to peak amplitude of the signal does not
exceed 4.4V.
Selected if positive or negative peak amplitudes are greater than
±5.1V or the peak to peak amplitude of the signal exceeds 5.1V.
±30mV ± 1% of trigger level reading.
±300mV ± 1% of trigger level reading.
-4
Hz to 200MHz (Prescaled by 2)
-4
Hz to 100MHz
D = Number of digits selected.
F = frequency rounded to nearest decade.
User Manual 2461 Type C