Table 6-2, Input 1 Sensitivity High Frequency Performance Limits - Racal Instruments 2461 User Manual

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Frequency
10Hz
5kHz
10kHz

Table 6-2, Input 1 Sensitivity high Frequency Performance Limits

Frequency
200MHz
160MHz
100kHz
10MHz
100kHz
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Signal level
25mV rms
25mV rms
25mV rms
Set the UCT measurement resolution to the values given in the
following table and apply the Frequencies given to the UCT
input 1. Check that the UCT gives a stable reading within the
tolerances given, at less than the signal levels given.
Signal level
50mV rms
50mV rms
50mV rms
25mV rms
25mV rms
Disconnect the test equipment.
Performance Verification and Adjustments 6-3
Resolution
Command
Syntax
SENS:RES 8
SENS:RES 8
SENS:RES 8
Resolution
Command
Syntax
SENS:RES 9
SENS:RES 9
SENS:RES 9
SENS:RES 8
SENS:RES 8
Tolerance
± 1Hz
± 0.1Hz
± 0.1Hz
Tolerance
± 1Hz
± 1Hz
± 1Hz
± 1Hz
± 0.1Hz

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents