200mhz universal counter timer type c
Table of Contents

Advertisement

Programming The Type C Function Card 4-52
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[:SENSe[1 | 2 | 3]]:FUNCtion
[:SENSe[1 | 3]]:FUNCtion
[:SENSe[1 | 2]]:FUNCtion
[:SENSe[1 | 2]]:FUNCtion
[:SENSe[1 | 2]]:FUNCtion
[:SENSe[1 | 2]]:FUNCtion
[:SENSe[1 | 2]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe[1]]:FUNCtion
[:SENSe]:FUNCtion
Description:
The SENSe:FUNCtion command selects a function and input
channel without changing most of the settings of the counter.
Initialization:
*RST
Frequency measurement function selected.
Example:
"SENSe1:FUNCtion
FREQuency "
Selects the frequency measurement function on channel 1.
Interrogative form:
[:SENSe[1 | 2]]:FUNCtion? ==> <Function String> (string in
short form).
Returns a string describing the currently selected function on
the specified channel. The possible strings are:
"FREQ"
"FREQ:RAT"
"VOLT:MAX"
"VOLT:MIN"
"VOLT:MIDD"
"PWID"
"NWID"
"PER"
"RTIM"
"FTIM"
2461Type C User Manual
"FREQuency "
"FREQuency:RATio"
"PERiod "
"VOLTage:MAXimum"
"VOLTage:MINimum"
"VOLTage:MIDDle"
"TINTerval"
"RTIMe"
"RISE:TIME"
"FTIMe"
"FALL:TIME"
"TOTalize"
"PWIDth"
"NWIDth"
"PHASe"
"CHECk"
Racal Instruments  1998