Input 1 Attenuator
Check

Figure 6-9, Input 1 X10 Attenuator Check Configuration

Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Set the UCT to its default state. Select 50
1.
Configure UCT as follows:
*RST
INP1:IMP 50
Connect the UCT to the signal generator as shown in Figure 6-
9.
SIGNAL GENERATOR
RF OUT
Set the signal generator to 1MHz, 10mV. Increase the signal
generator output level until the UCT gives a stable reading. Note
the signal generator output level.
Enable the UCT Input 1 attenuator and check that the UCT no
longer counts. Increase the signal generator until the UCT gives
a stable reading. Note.the signal generator output level and
check that this level is approximately ten times the level
previously noted.
Disconnect the test equipment.
returns unit to default state
select 50Ω impedance for Input 1