Racal Instruments 2461 User Manual page 102

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
FETCh? ---> 155
SENSe2:TOTalize:MANual OFF
FETCh? ->>235
SENSe2:TOTalize:MANual ON Re-starts the counting
FETCh? ---> 275
SENSe2:TOTalize:MANual OFF
FETCh? ->>395
SENSe:TOTalize:GATE OFF Resets the count
Interrogative form:
SENSe1:TOTalize:MANual? ==> <ON | OFF>
Returns the current state of the Manual Totalize
function.
*******************************************************
[SENSe]:RESolution <value | MINimum | maximum |
DEFault>
Description:
Selects the number of digits of resolution. This is in the range
of 3 to 10.
In frequency, period, ratio and check modes, the gate time is set
by the resolution selected. The resolution of phase and totalize
measurements is determined by the input signal.
The resolution setting may be changed indirectly by the
CONFigure:ARRay and MEASure:ARRay commands and by
the SENSe:APERture command.
Initialization:
*RST
Parameters:
3 ≤ value ≤ 10
MIN = 3
MAX = 10
DEF = 8
Programming The Type C Function Card 4-59
Returns an intermediate
count
Stops the counting
Returns the final count
Returns an intermediate
count
Stops the counting
Returns the final count