High Performance Trigger; Voltage Measurements; Individual Channel Filtering; Figure 1-3, High Performance Trigger - Racal Instruments 2461 User Manual

200mhz universal counter timer type c
Table of Contents

Advertisement

2461Type C User Manual
Time Out
High
Performance
Trigger
Voltage
Measurements
Individual
Channel Filtering
Racal Instruments  1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
system performance when of the input signal is missing.
In Manual mode, the trigger level is programmable from – 5.1 V to
+ 5.1 V with the exceptional resolution of 2.5 mV.
An automatic trigger mode is also available covering a frequency
range of DC and 50 Hz to 200 MHz.

Figure 1-3, High Performance Trigger

Automatic triggering is used to establish the peak DC voltages for
setting trigger points. This feature is used to measure Voltage
High, Voltage Low and Average Voltage.
The 2461 Type C card offers independent 50 kHz low pass filters
on each channel to allow measurements in noisy configurations.
General 1-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents