Safety Instrumented Systems (SIS) requirements
February 2020
8.5

Installation

The Rosemount 5408:SIS must be installed and configured as described in
installation
standard installation practices outlined in this manual.
The loop should be designed so the terminal voltage is within the limits specified in section
Power
Check that environmental conditions do not exceed the ratings in
reference
Note
The Rosemount 5408:SIS Level Transmitter is not safety-rated during maintenance work,
configuration changes, multidrop, loop test, proof-test, or other activity that affects the
safety function. Alternative means should be used to ensure process safety during such
activities.
8.5.1
Measuring range
The maximum measuring range is 82 ft. (25 m) for the Rosemount 5408:SIS Level
Transmitter in Safety (SIS) mode. Refer to
performance specification data.
8.6

Configuration

Use a HART-compliant master, such as Rosemount Radar Master Plus, AMS Device
Manager, or a Field Communicator, to communicate with and verify configuration of the
Rosemount 5408:SIS.
8.6.1
Prerequisites
Before doing any configuration, do the following:
Procedure
1. Write down the serial number from the transmitter label (see
2. Make sure you are connected to the correct transmitter by verifying the same serial
8.6.2
Configure device using Guided Setup
Follow the Guided Setup wizard for transmitter configuration, refer to Configuration.
When configuring parameters not included in the Guided Setup, it may be necessary to do
additional verification.
152
and
Electrical
installation. No special installation is required in addition to the
supply.
data.
number in your configuration tool.
Rosemount Radar Master Plus:
Under Overview, select Device Information → Identification.
AMS Device Manager and Field Communicator:
Select Overview → Device Information → Identification.