Suppress false echoes using AMS Device Manager
Procedure
1. Select Service Tools → Echo Tuning → Suppress. Suppressed echoes are shown in
2. Select Suppress or Remove Suppression.
3. Type the distance to the echo that should be added to or removed from the list, and
4. Select Finish when the Method Complete message appears.
Suppress false echoes using handheld communicator
Procedure
1. Select Service Tools → Echo Tuning → Suppress.
2. To view the current suppressed echoes, select Suppressed Echoes.
3. Select Suppress or Remove Suppression.
4. Type the distance to the echo that should be added to or removed from the list, and
7.4.3
Perform an analog loop test
Perform an analog loop test using Rosemount Radar Master
Plus
During a loop test, the transmitter outputs a fixed value (4 mA, 20 mA, or user-selected
value). The loop test command verifies the output of the transmitter, the integrity of the
loop, and the operations of any recorders or similar devices installed in the loop. A loop
test can also be used to determine the need for an analog output calibration.
Prerequisites
If the transmitter is equipped with a TEST terminal, current can be measured directly at the
terminal block without disconnecting any signal wires.
Procedure
1. Under Service Tools, select Simulate.
2. Select Loop Test.
3. Select 4 mA, 20 mA, or Other.
4. If Other is selected, enter the desired value.
5. Select Start.
6. Measure the loop current.
7. Select Stop to end loop test.
Related information
Calibrate analog out using Rosemount Radar Master Plus
Rosemount 5408 and 5408:SIS Level Transmitters
the table.
then select Next.
then select Enter.
Service and troubleshooting
February 2020
123

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 5408:sis

Table of Contents