Service and troubleshooting
February 2020
2. In the echo curve, click
Suppress a false echo manually
The false echo may also be suppressed manually if the position of the false echo is known.
Procedure
1. In Rosemount Radar Master Plus, under Service Tools, select Echo Curve.
2. Select Options.
3. Select Suppress False Echo Manually. Suppressed echoes are shown in the table.
4. Select Save.
5. Select Back.
122
select Delete.
Figure 7-8: Delete False Echo Suppression
To add a new suppression, select Add, and then type the distance to the false
echo and the width of the false echo area.
To change a suppression, select the cell you want change and type the new
value.
To delete a suppression, select the row you want to delete, and then select
Delete.
at the left end of the false echo suppression, and then