Reference Manual
00809-0100-4408
Flanged tank connection
Prerequisites
If applicable, assemble the segmented cone antenna (see
antenna).
Procedure
1. Place a suitable gasket on the tank flange.
2. Place the customer supplied flange over the gasket.
3. Tighten the bolts and nuts with sufficient torque for the flange and gasket choice.
4. Apply anti-seize paste or PTFE tape on threads according to your site procedures.
Rosemount 5408 and 5408:SIS Level Transmitters
Gasket may be used as a sealant for adapters with 1½- or 2-in. BSPP (G) threads.
Mechanical installation
February 2020
Assemble the segmented cone