3.7.4
Mount the threaded version
Procedure
1. Remove the lock nut.
2. Mount the O-ring.
3. Mount the antenna adapter on flange/manhole cover. Ensure the antenna adapter
4. Remove the M20 adapter and visually inspect the O-rings for damage and dirt.
Rosemount 5408 and 5408:SIS Level Transmitters
fits tightly to the flange/manhole cover.
A. Ø 3.98 ± 0.02 in. (Ø 101 ± 0.6 mm) or G 3½-in.
B. Max. 0.59 in. (15 mm)