Service and troubleshooting
February 2020
7.7.2
Tri-Clamp
Disassembly procedures
Procedure
1. Insert a wide flathead screwdriver into the groove at the base of the PTFE sealing.
2. Gently wiggle the screwdriver back and forth.
3. Repeat
4. Carefully lift the PTFE sealing.
144
®
version
Note
Be careful not to scratch or depress the PTFE surfaces (facing the process).
Step 1-Step 2
at different positions until the PTFE sealing is loose.