U3022AE10
Cal Kit Operational Check
Verification Limits
Specifications for the Test Set are typical. System performance for the network
analyzer and test set are only characteristic and intended as non-warranted
information. Only a functional certificate is provided for the test set.
It is recommended that you return your instrument to Keysight Technologies for
servicing or repair if the test set and network analyzer performance exceed the
operational verification limits.
Table 16
Typical Verification Limit for Reflection Tracking
10 MHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 10 GHz
10 GHz to 15 GHz
15 GHz to 20 GHz
20 GHz to 26.5 GHz
If you suspect that your test set configuration is not operating properly, ensure that
all front RF jumper interconnect cables are correctly attached
Keysight U3022AE10 User's and Service Guide
Frequency
Standard
-3 dB
-4 dB
-5 dB
-7 dB
-8 dB
a.
Reflection Tracking takes into account Source Loss, Receiver Loss, Margin,
and PNA Mixer Cal.
a
(Ports 1-12)
N5232A/B
Option 001
Standard
+4 dB
+3 dB
+3 dB
+3 dB
0 dB
-
-
N5242A/B
Option 001
-3 dB
+3 dB
-5 dB
+2 dB
-5 dB
+2 dB
-7 dB
+2 dB
-8 dB
-3 dB
-12 dB
-
61
Need help?
Do you have a question about the U3022AE10 and is the answer not in the manual?