temperature measurement
Problem
No temperature
measurement possible
Implausible measured
results,
temperature measurement
unsuccessful
Temperature probe cannot
be pushed completely into
the probe holder
COVIDIEN
GENIUS
TM
®

measurement

Problem
3000
Temperature module of the
SmartBucket not activated
Special patient condition, such as
hypothermia
Rectal measurement performed with
blue temperature probe
Oral/axillary measurement performed
with red temperature probe
Patient activity before oral temperature
measurement:
• Physical exertion
• Eating/drinking
• Brushing teeth
• Smoking
Set measuring position does not match
actual measuring position
No probe cover used
Temperature probe faulty
Probe cover not discarded
2 temperature
Patient temperature exceeds
measuring range of in-ear thermometer
Patient temperature undershoots
measuring range of in-ear thermometer
Ambient temperature exceeds
permitted range
Ambient temperature undershoots
permitted range
Cause
Cause
Remedy
Activate temperature module
(administrator rights required):
Retrofitting the SmartBucket
(seca mBCA 525 only)
• Assess patient vital signs using alternative
means
• Switch from predictive measurement
mode to direct mode
Only perform rectal measurement with red
temperature probe
Only perform oral/axillary measurement
with blue temperature probe
Only perform oral temperature
measurement about 20 minutes after any
of these activities
• Select measuring position according to
the probe in use
• Set correct measuring position on device
• Desinfect probe/ear thermometer
Disinfecting
• Use probe cover
Dispose of temperature probe, use
replacement
• Carefully withdraw temperature probe and
probe cover from the probe holder
• Discard probe cover
• Push the temperature probe into the
probe cover
Remedy
Assess patient vital signs using alternative
means
• Modify ambient temperature
• Perform measurement at a location with
an ambient temperature within the
permitted range
What do I do if ...? • 89