Spo 2 Measurement - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Problem
Measurement does not
start
Measured result
unexpectedly high
Measured result
unexpectedly low
Implausible measured
results
Different temperature units
on monitor and display
SpO
measurement
2
Fault
Implausible measuring
results
Measurement not possible
90 •
/
seca mVSA 535
seca mBCA 525
Cause
No probe cover used
Probe cover not properly located on
measuring head
Probe cover damaged
Temperature module of the
SmartBucket not activated
Probe cover damaged
• Lens of measuring head blocked
• Opening in probe cover blocked
Patient's auditory canal blocked
Probe cover damaged
Measuring position on in-ear
thermometer in wrong place
Ear thermometer faulty
Setting of unit on monitor and on in-ear
thermometer is not synchronized
automatically. If necessary, the monitor
converts measured results
automatically.
Cause
Intra aortic balloon pump affecting
pulse rate
Sensor damp
Sensor not applied correctly
Strong ambient light
Electromagnetic interference
Poor perfusion
®
only:
Masimo SET
Sensor service life expired
Sensor or patient cable defective
Remedy
• Desinfect probe/ear thermometer
Disinfecting
• Use probe cover
Ensure that probe cover engages audibly
with the measuring head
Dispose of damaged probe cover, use new
one
Activate temperature module
(administrator rights required):
Retrofitting the in-ear thermometer
(seca mVSA 535 only)
Dispose of damaged probe cover and use
new one
• Clean measuring head
• Dispose of probe cover, use new one
Clean auditory canal
Dispose of damaged probe cover and use
new one
Correct setting on in-ear thermometer (see
Operating Instructions for in-ear
thermometer)
Dispose of in-ear thermometer, use
replacement
• Press the °C/°F button on the in-ear
thermometer
• Change units on the monitor
(administrator rights required).
Remedy
Check pulse rate using ECG
• Dry sensor
• Use dry sensor
Apply sensor correctly, see instructions for
use for sensor
Cover application site with opaque material
• Switch off devices in proximity, isolate
interfering device
• Align interfering device differently or set up
in a different location
• Increase the distance between it and the
interfering device
• Assess patient vital signs using alternative
means
• Apply blood pressure cuff to location with
better perfusion
®
Use new Masimo SET
sensor.
Dispose of sensor or patient cable, use
spare part