Disinfecting - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Component (depending on version)
Temperature probe (red/blue) with cable
Probe holder (red/blue)
In-ear thermometer with cable
Magazine holder for probe covers (in-ear
thermometer)
SpO
sensor with cable
2
Patient cable for SpO

6.2 Disinfecting

Component (depending on equipment)
Monitor with
seca mBCA 525
compartment
Monitor with SmartBucket
seca mVSA 535
Measuring mat and electrode cables
Blood pressure cuff and compressed air
tube
Temperature probes (red/blue) with cable
Interval
As required
As required
As required
As required
Follow the manufacturer's instructions for use
sensor
Follow the manufacturer's instructions for use
2
1. Follow the instructions for use for the disinfectant.
2. Disinfect the device and its accessories as described in the table:
Interval
storage
As required
Before
and after a
measure-
ment
As required
As required
Cleaning
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with a mild soap solution
3. Clean temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
1. ➔
Removing the probe holder
2. Moisten a cotton bud with a mild soap solution
3. Wipe over all the surfaces of the probe holder
1. Do not discard probe cover
2. Moisten a soft cloth with a mild soap solution
3. Wipe over in-ear thermometer and cable
4. Allow to air-dry for approx. 30 minutes
5. Discard probe cover
1. ➔
Removing the magazine holder
2. Moisten a soft cloth or cotton bud with a mild soap
solution
3. Wipe over the surfaces of the magazine holder
4. Allow to air-dry for approx. 30 minutes
Disinfecting
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version):
Removing the probe holder
Removing the magazine holder
2. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
2. Wipe over measuring mat and electrode cables
3. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
2. Wipe over cuff and compressed air tube
3. Rinse thoroughly with water
4. Allow to dry at room temperature
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
3. Wipe over temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
English
Hygiene treatment • 81