Bioimpedance Measurement - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Fault
One or more LEDs on the
measuring mat not on
seca patient files cannot
be transmitted to the
measuring mat
No WiFi connection

9.3 Bioimpedance measurement

Fault
Results of bioimpedance
measurement deviate
significantly from expected
results
Some analysis modules
are not shown in the
"analysis" tab
Value of an analysis
parameter is shown in red
Cause
Measuring mat faulty
No WiFi connection
Infrared interface faulty
Distance between measuring mat and
monitor is too big
Cause
Patient position set on device does not
match actual patient position
Patient moved during the
measurement
The patient's arms or legs are not
spread away from the body
Electrode cables incorrectly assigned
Electrodes defective
Electrode cables or measuring mat
faulty
Incorrect seca patient file called up
List of analysis modules which can be
displayed has been restricted by
administrator
Value outside the normal range
determined for the analysis parameter
Remedy
Replace measuring mat
• Suspend the measuring mat in the stor-
age compartment of the device
• Call up seca patient file again, data will be
transmitted via the infrared interface
Inform seca Service
• Reduce distance
• ➔
Performing a measurement (without
WiFi direct connection to the measuring
mat)
Remedy
Ensure that the patient position set on the
device and the actual position match
Request the patient not to move during the
measurement and repeat the
measurement
Ask the patient to spread their arms and
legs away from the body
Ensure that the electrode cables are
connected to the correct electrodes to suit
patient position
Replace electrodes
Replace measuring mat
Transmit measurement to
and assign to the
seca analytics 115
correct seca patient file there
Administrator: Modify the analysis modules
which can be displayed
• Repeat the measurement to rule out
measurement error
• If the value remains outside the normal
range on a repeat measurement, take the
measured value into account as the
examination continues
What do I do if ...? • 87