3.4 Setting units of measurement

118 •
seca mVSA 535
/
seca mBCA 525
4. Confirm that you wish to proceed.
The calibration display appears.
5. Press the
symbol.
The symbol changes position.
6. Press the
symbol again.
The symbol changes its position again.
7. Repeat Step 6. until you get a request to confirm calibration.
8. Confirm calibration.
The touchscreen display is calibrated.
NOTE
If calibration is not confirmed, the procedure re-starts after a few
seconds.
CAUTION!
Patient hazard
In order to avoid misinterpretations, measuring results for medical use
must be displayed and used in SI units (weight: Kilograms/grams,
length: Meters/centimeters) only. Some devices offer the ability to dis-
play measuring results in other units. This is only an additional func-
tion.
► Use the results exclusively in SI units.
► The use of measuring results in non-SI units is the sole responsibility
of the user.