Sterilizing; Removing/Fitting Probe Holder (Devices With Temperature Probe); Removing The Probe Holder - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Component (depending on equipment)
Probe holder (red/blue)
In-ear thermometer with cable
Magazine holder for probe covers (in-ear
thermometer)
SpO
sensor with cable
2
Patient cable for SpO

6.3 Sterilizing

6.4 Removing/fitting probe holder (devices with temperature probe)

Removing the probe holder

82 •
/
seca mVSA 535
seca mBCA 525
Interval
As required
As required
As required
Follow the manufacturer's instructions for use
sensor
Follow the manufacturer's instructions for use
2
The device and accessories may not be sterilized.
1. Open the covering cap.
2. Remove the probe holder.
1. ➔
Removing the probe holder
2. Moisten a cotton bud with disinfectant (active ingredi-
ent: 70 % isopropanol)
3. Wipe over the inner surfaces of the probe holder
4. Allow to air-dry for approx. 30 minutes
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
3. Wipe over in-ear thermometer and cable
4. Wipe over probe tip, ensuring that all foreign particles
are removed.
5. Wipe probe tip dry with lint-free cloth (e.g. spectacles
cleaning cloth)
6. Allow to air-dry for approx. 30 minutes
1. ➔
Removing the magazine holder
2. Moisten a cotton bud with disinfectant (active ingredi-
ent: quaternary ammonium compounds)
3. Wipe over surfaces
4. Allow to air-dry for approx. 30 minutes
WARNING!
Patient hazard
The color of the probe holder indicates whether a device is intended
for oral/axillary or rectal temperature measurement. This distinction
can no longer be made once the probe holder has been removed.
Confusing the probe holders can lead to cross-contamination.
► Ensure that the probe holder is fitted back in the device from which
it was removed following a hygiene treatment.
Press!
Disinfecting