60 •
/
seca mVSA 535
seca mBCA 525
NOTE
You can only perform a further temperature measurement if you dis-
card the probe cover and push the temperature probe completely
back into the probe holder.
5. Push the temperature probe completely into the probe holder.
You have the following options for continuing:
► Measure further vital signs ➔
► Measure bioimpedance ➔
► Stop the measurement procedure ➔
procedure
Modifying presets
NOTE
Your settings apply to the current measurement procedure. If you
stop the measurement procedure ➔
dure, the presets defined by the administrator become active again.
1. Ensure that the temperature probe is pushed completely into the probe
holder.
2. Press the TEMP field.
The TEMP settings dialog window opens.
The presets are displayed.
3. Press the desired measurement method.
• Predictive
• Direct
4. Press the desired measuring position.
• Oral
• Axillary
5. Press the confirm button.
The dialog window closes.
Changed settings are used for the current measurement.
6. Perform a temperature measurement as described in the section entitled
Starting an oral/axillary temperature
=>
Measuring vital signs
Measuring bioimpedance
Stopping a measurement
Stopping a measurement proce-
measurement.