56 •
/
seca mVSA 535
seca mBCA 525
3. Press the
button on the monitor.
Blood pressure measurement starts:
Current cuff pressure is displayed.
The symbols for measurement procedure and measuring mode (in this
case: Single measurement, upward measurement) flash.
The measurement is automatically stopped as soon as valid blood
pressure values are detected.
The values for systolic/diastolic blood pressure SYS/DIA and mean
arterial pressure MAP are displayed.
You have the following options for continuing:
► Measure further vital signs ➔
► Measure bioimpedance ➔
► Stop the measurement procedure ➔
procedure
NOTE
• Using the Start button, you can cancel and restart blood pressure
measurement at any time.
• If an upward measurement does not deliver a measured value, the
device automatically switches to a downward measurement.
• If necessary, the device re-pumps several times during the down-
ward measurement (re-pumping: Increasing cuff pressure by
approx. 50 mmHg and releasing it in increments). If no measured
value is delivered even after several re-pumping steps, the process
is canceled and an error message is issued.
Measuring vital signs
Measuring bioimpedance
Stopping a measurement