66 •
/
seca mVSA 535
seca mBCA 525
6. Measure the patient's temperature:
a) Press the
b) Wait until you hear a triple acoustic signal
c) Take the measuring head out of the patient's ear
The monitor displays the temperature value.
The in-ear thermometer display likewise displays the temperature value
and the "Discard probe cover" symbol.
The temperature value is displayed on the monitor until you stop the
measurement procedure ➔
NOTE
The monitor always displays measured values in the unit set on the
monitor. If required, the values sent by the in-ear thermometer are
converted automatically.
7. Press the
button to discard the probe cover.
8. Dispose of the probe cover in line with your institution's policy.
9. Press the in-ear thermometer into the SmartBucket holder until you feel it
engage.
You have the following options for continuing:
► Measure further vital signs ➔
► Measure bioimpedance ➔
► Stop the measurement procedure ➔
procedure
button on the in-ear thermometer
Stopping a measurement
=>
Measuring vital signs
Measuring bioimpedance
Stopping a measurement
procedure.
Press!