9.4 Vital signs measurement

General

Problem
vital signs tab is not
displayed
The fields "Weight" and
"Height" are not displayed
in the vital signs tab
is being
seca mVSA 535
operated on a wheeled
stand: Measured results
for vital signs are
implausible

Blood pressure measurement

Problem
Implausible measured
results
Insufficient cuff pressure
88 •
/
seca mVSA 535
seca mBCA 525
General
Blood pressure measurement
COVIDIEN
COVIDIEN
SpO
measurement
2
Cause
USB connecting cable of the
SmartBucket not connected to the
monitor
SmartBucket not activated
module of the
seca 360° wireless
mVSA not activated
• Original seca USB connecting cable
not being used
• USB extension cable also in use
Cause
Excessive patient movement
Incorrect blood pressure cuff used
Blood pressure cuff not applied
correctly
Blood pressure cuff applied to an
extremity to which there is intravenous
access
Incorrect blood pressure cuff used
Blood pressure cuff or compressed air
tube leaking
Pump in device faulty
TM
TM
FILAC
3000 temperature measurement
TM
®
GENIUS
2 temperature measurement
• Connect the USB connecting cable of the
SmartBucket to the monitor
• Restart the device
Administrator:
Activate
SmartBucket:
system components
menu
Administrator:
Activate the
the
mVSA:
Administrator menu
• Use original seca USB connecting cable
(in scope of supply for wheeled stand)
• Do not use a USB extension cable
• Ask the patient to move as little as
possible
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Correctly apply blood pressure cuff, see
Instructions for Use for blood pressure cuff
Apply blood pressure cuff to a different
extremity
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Dispose of blood pressure cuff, use
replacement
Do not continue using device and have it
repaired by seca Service
Remedy
Administer
in the Administrator
module of
seca 360° wireless
Set up peripherals
in the
Remedy