Seca mVSA 535 Instructions For Use Manual page 99

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

LED wavelength:
Red
Infrared
Maximum light output
This information may be of particular interest to clinicians
Measuring time:
• Device switched on, sensor not applied
• Device switched off, sensor applied
Measuring range:
• SpO
2
• Pulse rate
• Perfusion index
Measuring accuracy:
• Measuring range
• SpO
(no patient movement)
2
• SpO
(patient movement)
2
• Pulse rate (no patient movement)
• Pulse rate (patient movement)
Low perfusion performance
• Pulse amplitude
• Transmission
• Oxygen saturation (SpO
• Pulse rate
Resolution:
• Oxygen saturation (SpO
• Pulse rate
IEC 60601-1: Medical electrical device, type BF
a.The accuracy of the Masimo SET
with induced hypoxia on healthy adult male and female volunteers with light to dark skin pigmentation in the range from 70 -
100% for SpO
compared to a laboratory CO oximeter and ECG monitor. This fluctuation corresponds to ± 1 standard
2
deviation. Plus or minus one standard deviation covers 68 % of the population.
b.The accuracy of the Masimo SET
with induced hypoxia on healthy adult male and female volunteers with light to dark skin pigmentation, 2 to 4 Hz rubbing and
tapping movements being performed at an amplitude of 1 to 2 cm and a non-repetitive movement being performed at 1 to
5 Hz at an amplitude of 2 to 3 cm in the range from 70 - 100% for SpO
monitor. This fluctuation corresponds to ±1 standard deviation and thus covers 68 % of the population.
c.The accuracy of the Masimo SET
Biotek Index 2™ simulator and the Masimo simulator with signal strengths of over 0.02 % and a transmission of over 5% for
saturations in the range from 70 to 100%. This fluctuation corresponds to ± 1 standard deviation. Plus or minus one standard
deviation covers 68% of the population.
d.The accuracy of the pulse rate of the Masimo SET
range 25 -240 min-1 compared to a Biotek Index 2™ simulator. This fluctuation corresponds to ± 1 standard deviation. Plus
or minus one standard deviation covers 68 % of the population.
e.The precise data can be found in the directions for use (DFU) for the sensors. Unless otherwise stated, the measuring position
for reusable sensors should be changed at least every 4 hours and for adhesive sensors, at least every 8 hours.
f.The sensor accuracy quoted applies in conjunction with the Masimo technology with a Masimo patient cable used for LNOP
sensors, RD SET sensors, the LNCS sensors or the M-LNCS sensors. The numbers stand for ARMS (accuracy root mean
square). Because pulse oximeter measurements are statistically distributed measuring procedures, only about two-thirds of
the measurements are expected to be in a range of ± Arms compared to the reference value. Unless stated otherwise,
accuracy for SpO
of 70% to 100% is quoted. The accuracy of the pulse rate is quoted from 25 to 240 min-1.
2
g. Masimo M-LNCS, LNOP, RD SET, and LNCS sensor types have the same optical and electrical properties and may differ
only in application type (adhesive/non-adhesive/hook & loop), cable lengths, optical component locations (top or bottom of
sensor as aligned with cable), adhesive material type/size, and connector type (LNOP 8-pin modular plug, RD 15-pin modular
plug, LNCS 9-pin, cable-based, and M-LNCS 15-pin, cable-based). All sensor accuracy information and sensor application
instructions are provided with the associated sensor directions for use.
h.Digit: Numerical value by which the last place of a value displayed can deviate from the actual measured value; used to state
the accuracy of a measuring device (example: displayed SpO
and 72%).
Masimo SET
®
)
2
)
2
®
technology with Masimo sensors was validated with no movement in human blood studies
®
technology with Masimo sensors was validated with movement in human blood studies
®
technology in the case of poor perfusion was validated on the test bench compared to a
®
technology with Masimo sensors was validated on the test bench for the
a b c d e f g
SpO
module
2
compared to a laboratory CO oximeter and ECG
2
value 70%, accuracy ± 2 digits; actual value is between 68%
2
660 nm
905 nm
15 mW
≤ 12 sec
≤ 8 sec
0 % - 100 %
-1
25 min
- 240 min
0.02 %-20 %
70 % - 100 %
70 % – 100 % ± 2 digits
0 % – 69 % not specified
70 % – 100 % ± 3 digits
0 % - 69 % not specified
-1
-1
25 min
- 240 min
± 3 digits
-1
-1
25 min
- 240 min
± 5 digits
> 0.02 %
> 5 %
± 2 digits
± 3 digits
1 min
Technical data • 99
English
-1
h
1 %