Spo Measurement Seca - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

SpO
measurement seca
2
Measurement
Measuring method
LED wavelength:
Red
Infrared
Maximum light output
This information may be of particular interest to clinicians
Measuring range:
• SpO
2
• Pulse frequency (default)
• Pulse frequency (enhanced)
Measuring accuracy:
• SpO
(no patient movement)
2
• SpO
(patient movement)
2
• Pulse rate (no patient movement)
• Pulse rate (patient movement)
Resolution:
• Oxygen saturation (SpO
• Pulse rate
EN 60601-1: Medical electrical device, type BF
a.Validated by clinical tests, where the measured values of the sensors were compared with those of the arterial co-oximetry
in healthy adult subjects over the specified functional oxygen saturation range.
b.As inherent to their functional principle, pulse oximetry measurements are statistically distributed; therefore only about two-
thirds of the measurement data are expected to fall within ±A
co-oximeter.
c.Tested with all motion patterns Fluke Index II oximeter tester.
seca SpO
)
2
a b
module
2
(Accuracy root mean square) of the value measured by a
rms
Functional oxygen saturation
Spectrophotometry (red/infrared)
660 nm
905 nm
0% - 100 %
-1
30 min
- 240 min
-1
20 min
- 300 min
60% - 70% ± 3 A
60% - 100% ± 2 A
<60% not specified
70% - 100% ± 3 A
<70% not specified
≤ 2 min
Technical data • 101
English
5 mW
-1
-1
rms
rms
c
rms
-1
-
1%
1 min
-1