Saving Energy; Switching The Monitor To Standby; Switching Off - Seca mVSA 535 Instructions For Use Manual

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Saving energy

Switching the monitor to standby

Switching off

40 •
/
seca mVSA 535
seca mBCA 525
The device provides three energy-saving modes which are activated
automatically after 5, 10 and 20 minutes if no entries are made:
After 5 minutes, the device switches to standby:
• The LED of the ON/OFF button is green.
• The touchscreen display goes off.
• The measuring mat remains switched on.
• All entries are retained.
After you switch back on, you do not need to log in again. Operation can be
continued.
After 10 minutes, the device goes to sleep:
• The LED of the ON/OFF button is green.
• The touchscreen display switches off.
• Data which have not been saved are lost.
• The measuring mat - if there is one - switches off.
After you switch back on, you have to log in again. Entries have to be made
again.
After 20 minutes, the device behaves as follows:
• Power supply operation:
– The charging process for the rechargeable batteries for the device
and the measuring mat - if there is one - starts automatically.
– The device remains asleep.
• Rechargeable battery operation:
– The device shuts down.
– The device switches off.
NOTE
As the administrator, you can specify from what battery charge status
the device shuts down ➔
► Briefly press the ON/OFF button of the monitor.
The LED of the ON/OFF button is green.
The touchscreen display goes off.
The monitor is on standby ➔
WARNING!
Electric shock
The monitor cannot be de-energized by pressing the ON/OFF button.
► Always remove the power supply connector and remove the re-
chargeable battery (if present and technically feasible) if the device
needs to be de-energized, e.g. for the hygiene treatment.
1. Ensure that the measuring mat - if there is a measuring mat - is correctly
suspended in the magnetic catch of the monitor ➔
measuring mat in the magnetic
2. Keep the ON/OFF button of the monitor depressed until the touchscreen
display and the LED of the ON/OFF button go off.
The device is switched off.
The measuring mat switches off automatically.
Power
management.
Saving
energy.
Suspending the
catch.