Seca mVSA 535 Instructions For Use Manual page 145

Spot check vital signs analyzer
Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

4. Press the seca 360° button.
5. Press the set up button.
6. Press one of the three wireless groups.
The device searches for
on.
The device will suggest three wireless channels.
NOTICE!
Incorrect device assignment
Only one example of a device category (e.g. personal scale or
stadiometer) can be integrated in each wireless group.
► Take note of the Technical Data in the section of the instructions for
use entitled ➔
seca 360° wireless
NOTICE!
Data transmission interference
The proposed channel numbers can be edited. This function is
intended purely for special reception situations. Devices may be
assigned to the wrong wireless groups or unreliable wireless
connections may arise.
► Only change the channel numbers if the proposed channel numbers
do not result in a reliable wireless connection.
► Ensure that the new channel numbers are not being used for other
wireless groups.
► Ensure that the channel numbers are at least 30 apart from one
another.
7. Switch on all
seca 360° wireless
wireless group.
When devices are detected, a beep will be heard.
Detected devices will be displayed on the monitor.
8. Press the finish button once all
detected.
seca 360° wireless
devices which are switched
system.
devices you wish to incorporate in the
seca 360° wireless
Setting up peripherals • 145
English
devices have been