Preparing The Product For Use; Operational Readiness; Charging The Batteries; Folding The Back Support Up/Down - Otto Bock 490E163 Series Instructions For Use Manual

Hide thumbs Also See for 490E163 Series:
Table of Contents

Advertisement

Note regarding the tyres
If the power wheelchair is not moved for several days, permanent colour changes may develop where it comes
into contact with the ground. A suitable base should therefore be used when parking it for extended periods of
time.
Avoid unnecessary parking outdoors. Direct exposure to sunlight/UV radiation causes the tyres to age more
quickly. As a result, the tread surface hardens and corner pieces break out of the tread.
The tyres must be changed when the tread is less than 1 mm (0.04") to ensure safe driving behaviour.
The tyres should be replaced every 2 years regardless of wear and tear.
When power wheelchairs with PU tyres are parked for longer periods, the tyres may become deformed (flat
spots). This deformation will go away on its own over time while driving.

5 Preparing the product for use

5.1 Operational readiness

To establish operational readiness, the plug connection between the battery cable and controller has to be connec­
ted. The plug connection is below the back of the battery cover.
1

5.2 Charging the batteries

WARNING
Incorrect charging or use of the battery
Injuries due to fire or explosion, damage to the battery, short circuit
► Observe the safety notices in the instructions for use (user), section "Use" > "Batteries/charging process".
WARNING
Insufficient ventilation of the battery charger while charging
Burns due to the battery charger overheating/catching fire
► Make sure the battery charger cannot overheat during the charging process.
► Ensure that the cooling fins/ventilation slots on the back of the device are not covered.
CAUTION
Wearing of unsuitable work clothing
Chemical burns due to insufficient protection
► Wear suitable work clothing (for example, gloves and protective eyewear) when working on the batteries.
See the power wheelchair instructions for use (user) for all information regarding the charging process.
The power wheelchair must be handed over to the user with fully charged batteries.

5.3 Folding the back support up/down

After unpacking, the backrest must be folded up and attached.
See the power wheelchair instructions for use (user) for more detailed information.
490E163=*
Connecting the battery cable
1) Take the plug of the battery cable in your hand. The
plug is located between the motors, below the bat­
tery cover (see fig. 1, left).
2) Connect the plug to the plug connection on the
controller (see fig. 1, right).
Disconnecting the battery cable
1) Disconnect the plug of the battery cable from the
controller. To release the plug, push the snap-fit
(see fig. 1, item 1).
2) Let the battery cable hang down loosely (see fig. 1,
left).
Preparing the product for use
9

Advertisement

Table of Contents
loading