Replacing the leak sensor
Warning:
including inhalation, always wear chemical-resistant, powder-free
gloves and observe Good Laboratory Practice when you handle them.
See the Material Safety Data Sheets for the solvents you use.
Warning:
solvents, flush the detector with an appropriate miscible solvent prior to
performing this procedure.
Warning:
compounds that are toxic, wear clean, chemical-resistant,
powder-free gloves when handling the leak sensor. To avoid
spreading contamination, do not allow contaminated gloves to
come in contact with uncontaminated surfaces, and discard gloves
immediately on completing the procedure.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Leak sensor
To replace the leak sensor:
1.
Power-off the refractive index detector.
2.
Open the refractive index detector's door, gently pulling its right-hand
edge toward you.
To avoid the harmful effects of personal contact with solvents,
To avoid the harmful effects of personal contact with leaking
To avoid personal contamination with biohazards or
Replacing the leak sensor
October 13, 2014, 715003547 Rev. C