12. In the column manager information window, click Control > Reset CM,
to reset the CM-30S.
Replacing the leak sensor
Perform this procedure if you must replace the leak sensor in the CM-30S.
Warning:
biologically hazardous or toxic materials, always wear clean,
chemical-resistant, powder-free gloves when performing this
procedure.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
New leak sensor
To replace the CM-30S's leak sensor:
Notice:
of a system module, do not disconnect an electrical assembly from
the module while the module remains connected to the ac supply
source.
Follow this procedure to completely interrupt power to the module:
Afterward, wait 10 seconds before disconnecting any assembly
from the module.
1.
Power-off the CM-30S.
To avoid personal contamination from contact with
To avoid damaging the electronic components or circuitry
1.
Set the module's power switch to Off.
2.
Disconnect the module's electrical supply cord from the ac
source.
Replacing the leak sensor
October 7, 2014, 715004178 Rev. B
73